Microlit NERO is a range of ergonomic single channel variable volume micropipettes with state-of-the-art technology such as UniCal™ and Air™ that will ensure great precision and convenience for our lab users. These innovative and patented technologies have been developed by our in-house product design engineers. The product is highly recommended for pathology, microbiology, and diagnostic testing labs where precise small volume pipetting is critical.
What’s in the box?
- Microlit Micropipette
- 1-2 compatible micropipette tips
- Calibration Tool
- Micropipette holder
- Calibration certificate
- Instruction manual
Achieve precision with uAir™ technology
The design of Microlit NERO ensures minimum dead air space between the piston and the liquid in the tip. This reduces air compression and enables the user to achieve precision in practical lab conditions.
Calibrate Efficiently with UniCal™ Technology
We have introduced a unique calibration mechanism/tool which allows quick in-lab calibration in a single operation WITHOUT disengaging the digits from the plunger mechanism.
Eject Tips Easily with our Tip Ejector
An in-built tip ejector is designed to eject tips with optimal force.
Use Various Tips with Universal Tipcone
Our tip-cone is designed to be compatible with most internationally accepted tips, which enhances the compatibility of the instrument in the industry.
Adjust volume with perfection
A distinct click at every volume change ensures perfect volume setting and prevents any accidental volume change.
Color Coding for Easy Identification
Color-coding for different volume ranges enables easy identification during daily laboratory routine.
Model No. | Description | Vol. Range(uL) | Increment(uL) | Accuaracy | CV | ||
---|---|---|---|---|---|---|---|
+% | +uL | +% | +uL | ||||
NRO-2 | Micropipette NRO Series Vol. Range 0.2-2µL | 0.2-2.0ul | 0.01 | 2 | 0.04 | 1.2 | 0.024 |
NRO-10 | Micropipette NRO Series Vol. Range 0.5-10µL | 0.5-10ul | 0.05 | 1 | 0.1 | 0.5 | 0.05 |
NRO-20 | Micropipette NRO Series Vol. Range 2-20µL | 2-20ul | 0.1 | 0.8 | 0.16 | 0.4 | 0.08 |
NRO-50 | Micropipette NRO Series Vol. Range 5-50µL | 5-50ul | 0.2 | 0.8 | 0.4 | 0.4 | 0.2 |
NRO-100 | Micropipette NRO Series Vol. Range 10-100µL | 10-100ul | 0.5 | 0.6 | 0.6 | 0.2 | 0.2 |
NRO-200 | Micropipette NRO Series Vol. Range 20-200µL | 20-200ul | 1.0 | 0.6 | 1.2 | 0.2 | 0.4 |
NRO-1000 | Micropipette NRO Series Vol. Range 100-1000µL | 100-1000ul | 5.0 | 0.6 | 6 | 0.2 | 2 |
NRO-5000 | Micropipette NRO Series Vol. Range 500-5000µL | 0.5-5ml | 20.0 | 0.6 | 30 | 0.2 | 10 |
NRO-10000 | Micropipette NRO Series Vol. Range 1-10ml | 1-10ml | 50.0 | 0.6 | 60 | 0.2 | 20 |
The error limits (Accuracy and Coefficient of Variation) mentioned above are in accordance with the nominal capacity (or maximum volume) indicated on the instrument. These are obtained by using the instrument with distilled water at equilibrium, the ambient temperature of 20 °C while operating it smoothly and steadily. The error limits are in accordance with DIN EN ISO 8655-2.